UDI and medical supply traceability: what hospitals need to prepare

Article cover: UDI and medical supply traceability

MEDCONS Expert Team · Updated: 6 November 2026

Summary: UDI (Unique Device Identification) is a unique code that the manufacturer places on a product, made up of a part that identifies the model and a part with production information such as lot number and expiry date. Vietnam already has TCVN 13996:2024 on unique device identification for medical devices. To use UDI for traceability, hospitals need a standardised supply catalogue, software that can read and store the code, and a process for scanning at the points of receipt, issue and use.

What is UDI?

UDI (Unique Device Identification) is a unified identification system for medical devices and supplies. Under TCVN 13996:2024, the code has two parts:

  • UDI-DI (Device Identifier): the fixed part, identifying the manufacturer and the product model and packaging.
  • UDI-PI (Production Identifier): the part that changes with each lot or each item, containing information such as lot number, serial number and expiry date.

The code can be shown as a linear barcode, a two-dimensional DataMatrix code or an RFID tag. The standard was drafted by Technical Committee TCVN/TC/GS1 and announced by the Ministry of Science and Technology; its scope covers devices on the market and at healthcare facilities.

Worldwide, UDI is already mandatory in many markets such as the United States (FDA) and Europe, so most supplies imported from large manufacturers already carry this code on their packaging.

Why does traceability matter for hospitals?

At the workshop “Applying traceability in healthcare”, organised in 2021 by the National Barcode Center and the Department of Medical Equipment and Construction (Ministry of Health), speakers highlighted the main benefits:

  • Patient safety: when a lot is subject to a warning or recall, the hospital quickly identifies which store it is in and which patients it was used for.
  • Fighting counterfeit and substandard goods: a standard code helps verify origin right at goods receipt.
  • Management and planning: standardised data supports purchasing, tendering and payment.

The workshop also pointed out a common situation: many hospitals use several coding systems in parallel – manufacturers’ codes, codes for health insurance payment, self-assigned internal codes – which makes data hard to align across departments, systems and suppliers.

How does traceability work inside a hospital?

Imagine this situation: a manufacturer announces a recall of a lot of catheters. If the hospital has recorded the UDI at every step, it only needs to look up the lot number in the system to know at once: when the lot was received, from which supplier, how much is left in the main store, which departments it was issued to and which patients it was used for. The remaining stock is blocked from issue immediately. Without this data, the hospital has to check every store and every ledger by hand, which takes days and may still miss something.

What do hospitals need to prepare?

1. Standardise the supply catalogue

Each item needs a single record in the catalogue that can be linked to the manufacturer’s code (UDI-DI or GTIN), the internal code and the code used for payment. This is the most labour-intensive task, but it determines data quality later on.

2. Software that can read and store lot and expiry information

The supply management system must be able to separate the fixed and variable parts of the code and fill in the lot number and expiry date automatically on scanning, instead of manual entry. This information must follow the item through every step: goods receipt, store transfer, issue to departments and use for patients.

3. Suitable scanning devices

DataMatrix codes on supply packaging require a 2D scanner. For large stores or operating theatres, handheld devices (PDAs) can also be considered for scanning right at the shelf or at the point of use.

4. A scanning process at key points

Traceability is only complete when there are no “breaks”: goods receipt from suppliers, issue to departments and recording use for patients. Rules are needed on who scans, when they scan, and how to handle items without a code or with an unreadable code.

5. Coordination with suppliers

Hospitals should include requirements for barcodes, lot and expiry information in their supply conditions, and ask suppliers to send product data in a standard format to reduce the initial data-entry work.

6. Training and regular checks

Store staff and department nurses need to be shown how to do it. Regular stocktakes by scanning help detect discrepancies and near-expiry items early.

Suggested roadmap

  1. Assess the current catalogue, coding systems and store processes.
  2. Standardise the catalogue and link manufacturers’ codes, starting with high-value supply groups.
  3. Pilot scanning at the main store and one high-consumption department.
  4. Evaluate, adjust the processes and then roll out hospital-wide; connect with HIS and accounting.

Frequently asked questions

How is UDI different from barcodes the hospital prints itself?
Self-printed codes only have meaning inside one hospital. UDI is applied by the manufacturer according to a common standard, so the same product is identified in the same way by the manufacturer, the distributor, the hospital and the regulator.

How are supplies without UDI managed?
The hospital can still assign an internal code at goods receipt and record the lot number and expiry date from the documents. The software should allow both ways: scan the UDI when there is one, use the internal code when there is not.

Is scanning mandatory for every supply?
It is best to start with high-value supplies, implants and high-risk supplies, then expand gradually. Low-value, high-volume supplies can be managed by lot at store and department level.

Does UDI replace the code used for health insurance payment?
No. Each type of code serves a different purpose. What matters is that the hospital’s supply catalogue links these codes together in the same record.

Note: the validity and application requirements of standards and regulations may change; hospitals should check the original documents and the latest guidance from the authorities.

MEDCONS supports hospitals in standardising catalogues, designing processes and implementing the SupplyCore software, with lot and expiry management and code scanning. See also the service Healthcare digitalization and the article 5 trends in hospital consumables management.

References

  • TCVN 13996:2024 – Article numbering and barcoding – Unique device identification for medical devices (in Vietnamese): hethongphapluat.com
  • Thương hiệu & Công luận (2021), “Applying traceability in healthcare” (in Vietnamese): thuonghieucongluan.com.vn